4 calibration using an external weight – A&D Weighing FX-3000i User Manual

Page 29

Advertising
background image

27

7-4 Calibration Using an External Weight

This function calibrates the balance using an external weight.

Operation

1 Connect the AC adapter and warm up the balance

for at least 30 minutes with nothing on the pan.

2 Press and hold the CAL key until Calout is

displayed, and then release the key.

3 The balance displays Cal 0 .

If you want to change the calibration weight (a

list of usable weights is shown on page 24),
press the SAMPLE key and proceed to step 4.

If you use the calibration weight value stored in

the balance, proceed to step 5.

4 Specify the calibration weight value as follows:

SAMPLE key

To switch the display condition

to: “All of the segments
blinking” (calibration weight
selection mode) or “The last
two digits blinking” (value
adjustment mode).

RE-ZERO key (To increase the value)
MODE key (To decrease the value)

To select the calibration

weight or adjust the value.

PRINT

key

To store the new weight

value. Even if the AC
adapter is removed, the data
is maintained in non-volatile
memory.

CAL key

To cancel the operation and

return to Cal 0 .











Select a

weight

To next page

Press and

hold the key

Release

the key

e.g. Calibration

weight value
1000.12 g

Advertising
This manual is related to the following products: