2 initializing the balance – A&D Weighing FX-3000i User Manual

Page 34

Advertising
background image

32

Operation

1 Press the ON:OFF key to turn off the display.

2 While pressing and holding the PRINT key and the SAMPLE key, press the ON:OFF key.

The balance displays p5 .

3 Press the PRINT key. Then the balance displays the function switches.

4 Set the switches using the following keys.

SAMPLE key

To select the switch to change the parameter.

RE-ZERO key

To change the parameter of the switch selected.

0:To inhibit changes. 1:To permit changes

PRINT key

To store the new parameter and return to the weighing mode.

CAL key

To cancel the operation ( Clr is displayed.) To return to the

weighing mode, press the CAL key once again.

8-2 Initializing the Balance

This function returns the following parameters to factory settings.

Calibration data
Function table
The sample unit mass value (counting mode), 100% reference mass value (percent mode)
External calibration weight

Function switch settings

Statistical data

Internal mass value (Only for FZ-i series)


Note
Be sure to calibrate the balance after initialization.

Operation

1 Press the ON:OFF key to turn off the display.

2 While pressing and holding the PRINT key and the

SAMPLE key, press the ON:OFF key. The balance

displays p5 .


3 Press the SAMPLE key to display Clr .

4 Press the PRINT key.

To cancel this operation, press the CAL key.

5 Press the RE-ZERO key.

6 Press the PRINT key to initialize the balance.

The balance will automatically return to the weighing
mode.

With these keys

held down

Press the key