5-6. Weighing range for the EW-i series

‰

The EW-i series have three weighing ranges, and the display shows which range

the weight value belongs to with the mark R1, R2 or R3.

‰

There is a function setting to select how the weighing range changes.

‰

Select from automatic range (rng 1), manual range (rng 0) or fixed range (rng 2

to 4).

Operation

Function setting

Operation

rng 1

Automatic range

‰

When the weight value exceeds the maximum value of

the range, the weighing range changes automatically
from a lower to a higher weighing range.

‰

When there is nothing on the weighing pan and the

display shows zero with the ZERO mark, the weighing
range changes from a higher to the lowest range.

‰

When the RE-ZERO key is pressed in a higher range,

the balance will be tared to zero and the weighing range
changes to the lowest range.

rng 0

Manual range

‰

Press the SAMPLE key when the display shows a

weight value (neither counting nor % display). The
weighing range changes to a higher range at any load.

‰

Press the SAMPLE key to change from a higher to the

lowest range, when there is nothing on the weighing pan
and the display shows zero with the ZERO mark.

‰

When the RE-ZERO key is pressed in a higher range,

the balance will be tared to zero and the weighing
range changes to the lowest range. If the weight of the
object is not more than 2% of the weighing capacity, the

RE-ZERO key doesn’t tare, but zeroes the balance

and the weighing range doesn’t change.

Press the SAMPLE key to change the weighing range

at zero display.

rng 2 to 4

Fixed range

(

*

)

‰

The weighing range is fixed. Set the function to the

weighing range according to the purpose.

(

*

)

This function will not be available for some of legally certified models.