2. data format, 3. data output mode – A&D Weighing EW-12Ki User Manual

Page 25

Advertising
background image

23

9-2. Data format

S T

,

+

0

0

0

0

0

.

0

0

g C

R

L

F

Header

Data

Unit Terminator



‰

There are four types of headers:

ST : Stable weight data (including % data)

QT : Stable counting data

US : Unstable weight data (including count and %)

OL : Out of weighing range (Over)


‰

The data is normally 9 digits including a decimal point and a sign.

‰

There are 11 types of units:

g : Weight data “gram”

P C : Counting data “pcs”

% : Percentage data “%”

o z : Weight data “decimal ounce”

l b : Weight data “decimal pound”

o z t : Weight data “troy ounce”

c t : Weight data “carat”

m o m : Weight data “momme”

d w t : Weight data “penny weight”

G N : Weight data “grain”

t l : Weight data “tael”


‰

The terminator is always C

R

L

F

.


‰

Example of output data:

Weight data “gram”

S T

,

+ 0 0 1 2 3 4

.

5

g

C

R

L

F

Counting data

Q T

,

+ 0 0 0 1 2 3 4 5

P C

C

R

L

F

Percentage data

S T

,

+ 0 0 0 1 2 3

. 4

%

C

R

L

F

Out of range “gram” (+)

O L

,

+ 9 9 9 9 9 9

. 9

g

C

R

L

F

Out of range “pcs” (-)

O L

,

- 9 9 9 9 9 9 9 9

P C

C

R

L

F


9-3. Data output mode

Stream mode

Set the function “prt 0”.
The balance outputs the current display data. The data-update rate is approximately
10 times per second. This rate is the same as the display-update.
The balance does not output data while it is in the setting mode.

Separator

(“ ” shows a space.)

Advertising
This manual is related to the following products: