10. ID NUMBER AND GLP

The ID number is used to identify the balance when Good Laboratory Practice (GLP)
is used. The following GLP data is transmitted to an AD-8121 printer or a computer
using the RS-232C interface.

‰

The result of calibration (“Calibration report”)

‰

The result of calibration test (“Calibration test report”)

‰

The “Start block” and “End block” for GLP data


10-1. Setting the ID number

1. Press and hold the SAMPLE key to display

func .


2. Press the SAMPLE key several times to display

id .


3. Press the PRINT key. Enter the ID number using

the following keys.

SAMPLE key To select the digit blinking to

change.

RE-ZERO key To set the character of the digit

selected. See the table below for
the “display character set”.

PRINT key

To store the value and proceed to
the next step.

MODE key

To cancel the value and proceed
to the next step.


4. When the above operation has completed, func

appears after end .


5. Press the MODE key to return to the weighing

mode.



Display character set

0 1 2 3 4 5 6 7 8 9 - A B C D E F G H I J K L M N O P Q R S T U V W X Y Z
0 1 2 3 4 5 6 7 8 9 - a b C d e f g H i j k l m n o p q r s t U v w x y z

“ ” : Space



Press

several times

func

id

000000

end

func

Set using the

relevant keys

Returns to

the weighing mode

Press

several times

Press and hold