Comparator, Func poff 0 cp 0 cp 3, 1. setting example – A&D Weighing EW-12Ki User Manual

Page 14

Advertising
background image

12

6. COMPARATOR

The results of the comparison are indicated by HI, OK or LO on the display.
The comparison is as follows:
LO

< Lower limit value ≤ OK ≤ Upper limit value < HI

Operating conditions (see the function setting “Cp”):

‰

No comparison (comparator function disabled).

‰

Compares all data.

‰

Compares all stable data.

‰

Compares plus data except those near zero (plus data greater than +4d).

‰

Compares stable plus data except those near zero (stable plus data

greater than +4d).

‰

Compares all data except those near zero (all data greater than +4d or

less than -4d).

‰

Compares stable data except those near zero (stable data greater than

+4d or less than -4d).

d = the smallest display division

e.g.: 4d=four display divisions

The upper limit and lower limit numerical values are common to each of the weighing,
counting and percent mode. An example for the EK-1200i/2000i/3000i is as follows.

Upper limit value “001010”: “101.0g” “1010pcs” “101.0%”
Lower limit value “000990”: “99.0g” “990pcs” “99.0%”


6-1. Setting example

This example will be “Compares plus data except
those near zero”.

Selecting a comparison mode

1. Press and hold the SAMPLE key to display

func .

(If the comparison mode is already set, press the

SAMPLE key to go to “Entering the upper and lower limit values”.)


2. Press the PRINT key, then the balance displays

poff X .


3. Press the SAMPLE key several times to display

Cp X .


4. Press the RE-ZERO key several times to display

Cp 3 .


5. Press the PRINT key to store the settings.

Cp Hi appears after end .



func

poff 0

Cp 0

Cp 3

Press and hold

Press

several times

Press

several times

To store

Advertising