15

5. At step 3, pressing the PRINT key weighs the

zero-point value. Do not touch the pan during
weighing.
When the zero calibration is completed, the display
shows the calibration weight value.


‰

To perform the zero calibration only and finish the

procedure, press the MODE key.











6. Place the calibration weight with the same value as

displayed on the pan. Press the PRINT key to
weigh it. Do not touch the pan during weighing.







7. end appears.

Remove the weight from the pan, and press the CAL
switch or MODE key to return to the weighing
mode.

‰

Note

The value set in step 4 is stored in memory even after the power is

switched off.

If the balance is to be moved to other places, set the gravity

acceleration value of the area where the calibration using a weight is
to be done, and calibrate the balance according to the procedure
above. See the next section to set the value.










1000.0