CAL errors

Warning to indicate that calibration has been canceled
because the calibration weight is too heavy.

Warning to indicate that calibration has been canceled
because the calibration weight is too light.

Check the weighing pan and the calibration weight.
To return to the weighing mode, press the MODE key.

Low battery

Warning to indicate that the battery pack (OP-09) is
discharged. Immediately stop using it, and recharge or
use the AC adapter.

AC adapter error

Warning to indicate that the output voltage of the AC
adapter is too high. Check if the AC adapter is correct.


Stability error

Warning to indicate that the weight value is not stable
and the balance cannot display it. Prevent vibration and
drafts. Press the MODE key to return to the weighing
mode.

Internal errors (# = 2,3,4 or 6)

The balance detects error state in the internal processing.
With nothing on the weighing pan, turn the power off and
on again. If the error persists, request service.

If you cannot cancel an error or other errors have occurred, request service from
the store where you purchased the balance or from your local A&D dealer.