Calibration test report

The calibration test mode is used to compare a calibration test weight with the
calibration test data weighed by the balance.
‰

This test does not perform calibration and this mode is available even after sealing

the calibration (CAL) switch.



1. Press and hold the SAMPLE and PRINT

keys. CC will appear. Then, release both
keys.


‰

Pressing and holding the calibration (CAL) switch

will also display CC after Cal . Then,
release the CAL switch while showing CC .


2. CC 0 is displayed.

3. If necessary, change the value of calibration test

weight according to the procedure step 4 of “7-1.
Calibration using a weight”.


4. With nothing on the pan, press the PRINT key.

The zero point is measured and the weighed value
is displayed for a few seconds. Then, the display
shows the value of calibration test weight.




5. Place a weight of the same value as displayed on

the pan and press the PRINT key to weigh it.
The weighed value is displayed for a few seconds.








6. end appears.


7. glp is displayed and calibration test report is

output.



8. end appears again. Remove the weight and

press the MODE key to return to the weighing
mode.




CC

CC 0

Press and hold both keys.

Release both keys.

0.0

1000.0

1000.1

end

glp

Returns to

the weighing mode

end

To output calibration

test report