10-2. GLP report

‰

To print the GLP report to the AD-8121B, select the function setting “info 1” and

“pUse 1” for the balance, and use MODE 3 of the printer.

‰

To transmit the GLP report to a personal computer, select the function setting “info

2” and “pUse 0”.

‰

The serial number attached to the lateral side of the balance may have a form

“#Q1234567” (# = I, E and so on). In this case, the first letter “#” is ignored as a
GLP data and output data has a form “Q1234567”.

Calibration report

‰

This function is not available after sealing the calibration (CAL) switch.


1. Perform calibration according to “7-1. Calibration

using a weight”.



2. end appears when the calibration has

been completed.



3. glp is displayed and calibration report is

output.


4. end appears again. Remove the weight

and press the MODE key to return to the
weighing mode.


AD-8121 format “info 1”

General format “info 2”
















: Space,

ASCII

20h

CR: Carriage return, ASCII 0Dh
LF: Line feed, ASCII 0Dh

end

glp

To output

calibration report

Operation of Calibration

(See “7.CALIBRATION”)

end

Returns to the weighing mode

`~~~~~~~~~~

A

~

&

~

D<CRLF>

MODEL

~~~

EK-1200i<CRLF>

S/N

~~~~~

Q1234567<CRLF>

ID

~~~~~~~~

ABCDEF<CRLF>

DATE<CRLF>
<CRLF>
TIME<CRLF>
<CRLF>
CALIBRATED(EXT.)<CRLF>
CAL.WEIGHT<CRLF>

~~~~~~

+1000.0

~~

g

<CRLF>

SIGNATURE<CRLF>
<CRLF>
<CRLF>
---------------

~

<CRLF>

<CRLF>
<CRLF>

A & D

MODEL EK-1200i
S/N Q1234567
ID ABCDEF
DATE 2011/11/14
TIME 02:53:21
CALIBRATED(EXT.)
CAL.WEIGHT
+1000.0 g
SIGNATURE


- - - - - - - -

Manufacturer

Model

Serial number

ID number

Date

Time

Calibration

executed

Calibration weight

Column for

signature