Comparator – A&D Weighing SE-150KAL User Manual

Page 17

Advertising
background image

15

8. COMPARATOR

Compares the weight value with the upper and lower limit values and shows the results
of the comparison by

HI

OK

LO

on the display.

The formula to compare is as follows:

1

LO

< Lower limit value ≤

OK

≤ Upper limit value <

HI

Select comparison conditions that suit your own application. (Refer to “Comparator mode: Cp”
in “10-2. Function list”.)

0: No comparison (comparator function disabled).
1: To compare all data.
2: To compare all stable data.
3: To compare all data which are more than +4d or less than -4d.
4: To compare stable data which are more than +4d or less than -4d.
5: To compare all data which are more than +4d.
6: To compare stable data which are more than +4d.
d = minimum display (Refer to “13-1. Specifications”.)
Also in the counting mode, “d” is equal to the minimum display of kg mode.

‰

The upper and lower limit values are maintained even if the power is turned OFF.

‰

The upper and lower limit values are common to the weighing and counting mode.

‰

Ignore the decimal point when setting the upper limit or lower limit value.

For example, when the weighing capacity is 30 kg and the setting value is “001000”:

Display mode

Limit value

Capacity / Minimum display

Normal resolution kg

10.00 kg

30.00 kg / 0.01 kg

High resolution kg

1.000 kg

30.000 kg / 0.005 kg

Normal resolution lb

10.00 lb

66.00 lb / 0.02 lb

High resolution lb

10.00 lb

66.00 lb / 0.01 lb

Normal resolution oz

100.0 oz

1050.0 oz / 0.5 oz

High resolution oz

100.0 oz

1050.0 oz / 0.2 oz

Counting mode

1000 pcs



















Advertising
This manual is related to the following products: