Options, Maintenance, 2. error codes – A&D Weighing SE-150KAL User Manual

Page 24

Advertising
background image

22

11. OPTIONS

The following options are available for the SC/SE series:

‰

OP-02 (SCE-02)

USB interface

‰

OP-03 (SCE-03)

RS-232C serial interface and Comparator relay output

Note

When an option is installed, the dust-tight and water-tight performance of the
scale will be degraded.

For details, refer to the instruction manual for the options.

12. MAINTENANCE

12-1. Notes on maintenance

‰

Do not disassemble the scale. Contact your local A&D dealer if the scale needs

service or repair.

‰

Use the original packaging for transportation.

‰

Do not use organic solvents to clean the scale. Use a warm lint free cloth

dampened with a mild detergent.

‰

Calibrate the scale periodically to keep weighing accuracy.

12-2. Error codes

Overload error

Indicates that an object beyond the weighing capacity
has been placed on the weighing pan.
Remove the object from the weighing pan.

Underload error

Indicates that the weight sensor receives strong force
upward.
Check if the weighing pan is touching anything or if there
is anything in the base. There is a possibility that the
weight sensor or internal circuit may have a failure.

Unit weight error

Indicates that the sample weight is too light to set the unit
weight in the counting mode.
Increase the number of samples.

Low battery

Indicates that the batteries are depleted.
Replace them with new batteries.

Other

The internal circuit may have a failure.

(

indicates an error number.)


‰

If the error persists or other errors occur, contact your local A&D dealer.