10-2. Function list

Item

Parameter

Description

0

Auto power-off disabled

Auto power-off
function

poff

Š

1

Auto power-off enabled

Turns the power OFF
automatically

0

Normal (1/3,000)

Weight display
resolution

r5eo

Š

1

High (1/6,000 or 1/7,500)

Changes “d”,

minimum display

Š

0

2400 bps

1

4800 bps

Serial interface
Baud rate

bp5

2

9600 bps

Š

0

Stream mode

Command mode

1

Command mode only

2

Print switch mode

Command mode

3

Auto-print mode +/- data

Command mode

Serial interface
Data output
mode

prt

4

Auto-print mode + data

Command mode

0

No reply except “Q” command

Serial interface
Response

aCk

Š

1

Reply to commands

Š

0

Comparator disabled

1

Compares all data

2

Compares all stable data

3

Compares data of > +4d or < -4d

4

Compares stable data of > +4d or < -4d

5

Compares data of > +5d

Comparator
mode

Cp

6

Compares stable data of > +5d

Comparison conditions

d = minimum display

0

Weak / fast response

Š

1

Normal / normal response

sStability
Response speed

Cond

2

Strong / slow response

0

Zero tracking function disabled

Zero tracking

trc

Š

1

Zero tracking function enabled

Tracks the zero drift

Š

0

Dot (.)

Decimal point

dp

1

Comma (,)

Selects a decimal
point type

.

no Initialization disabled

Initialization

Clrf

go Initialization enabled

Restores the function
settings to the factory
setting values.

Š

Factory setting

d = minimum display, the minimum mass that can be weighed

Even the counting mode uses “d” for judgment.