Functions, 1. setting the parameters – A&D Weighing SE-150KAL User Manual

Page 22

Advertising
background image

20

10. FUNCTIONS

The scale has function settings to specify the scale performance.
The parameters set in the function settings are maintained even if the power is turned
OFF.





10-1. Setting the parameters

1. Turn the power OFF.

2. Press and hold the TARE switch and press the

ON/OFF switch to turn the power ON. The software

version is displayed.

‰

∗∗∗” indicates the software version number.

3. Press the UNITS switch to display the item.


4. Change the item or parameter using the following switches.

UNITS To display the next item.

PRINT To increase the value of the blinking

digit by one (to change the parameter).

‰

When the parameter is changed,

the STABLE indicator turns off.


5. Press the TARE switch to store the setting value.

After “

end

”, the scale returns to the weighing mode.



To restore the function settings to the factory
setting values

When “

Clrfno

” is displayed, press the PRINT switch to

change to “

Clrfgo

”. Then, press the TARE switch. The

function settings are restored to the factory setting values.
After “

end

”, the scale returns to the weighing mode.






Item

Parameter

While

holding down, press

Press

Press

Press

When changed

ON/OFF

Press

Press

Advertising
This manual is related to the following products: