5. calibration calibration calibration calibration – A&D Weighing AD-4407 User Manual

Page 13

Advertising
background image

AD-4407 Weighing Indicator

Page

11

5.

5.

5.

5. Calibration

Calibration

Calibration

Calibration

This indicator, converts an input voltage from a load cell to the "mass" value, and

displays it. Calibration is the adjustment function so that the scale (indicator) can

display the weight correctly.

5.1.1.

Items of Calibration Mode

There are four items in the calibration function in the procedure.

How to calibrate: In weighing mode, press the CAL key. After Cal in is displayed

for 2 seconds Cal 0 will appear. Then the required items should be selected and

displayed with the

MODE

key, then executed by pressing

ENTER

key.

NOTE: Calibration could be started by simultaneously pressing the

ZERO

and

TARE

key,

instead of CAL key. However, in some cases, the procedures are altered to nullify this

option. Furthermore, this option is not available with Stamping type version setting.

Required Items

Cal 5et

Store capacity, resolution, alignment of decimal point position and
display format, weighing range and unit. These items should be input
first in order for the indicator to function as a weighing instrument.
These parameters do not need to be changed again unless the
indicator itself is replaced. For details, refer to “5.2.1.Configuring a
Weighing Instrument”.

Cal 0

Calibrates zero and span. This is required after installation, to get
accurate data. For details, refer to “5.2.3. Zero Calibration” and “5.2.4.
Span Calibration”.

Optional Items (Sub-functions)

lnr 0

Performs digital linearization. Refer to “5.4.Digital Linearization
Function”.

g 5et

Compensates for acceleration of gravity.

Refer to “5.5. Gravity

Compensation Function”.

Gravity compensation function: Compensates for weighing error
between the calibration location and other weighing location using
gravity acceleration.

In the calibration mode the keys have functions as follows:

0

-

9

Numerical

keys.

MODE

The key to display other items.

CLEAR

The cancel key at inputting data, initial data and changing mode.

+/-

The key to display other parameters.

ESC

The key to proceed to the next step without changing them.

ENTER

The key to store new calibration data and proceeds to the next step.

CAL

The key to store all data into memory and display Caloff after the

calibration mode. And press the

ON/OFF

key to turn off the display.