A&D Weighing AD-4407 User Manual

Page 30

Advertising
background image

Page

28

AD-4407 Weighing Indicator

Current Loop Output

Item

Parameter

Description

** 0

Displayed data

1

Gross data

2

Net data

3

Tare data

f36

Output data

4

Gross data / Net data / Tare data

0

Stream mode

1

Manual mode

2

Auto print mode(+)

3

Auto print mode(+/-)

4

When accumulation, automatically output

f37
Output mode

** 5

Not used

** 0

No delay

f38
Delay for continual data

1

2.0 seconds (F36=4, except F37=0)

0

600 bps

1

1200 bps

f39
Baud rate

** 2

2400 bps

*: Initial settings.

bps:

bit per second.

Serial Interface

Item

Parameter

Description

** 0

Stream mode, command is not acceptable

1

Manual mode, command is effective

2

Auto print mode(+), command is effective

3

Auto print mode(+/-), command is effective

4

When accumulation, automatically output

f40
Output mode

5

Command mode (output by command only)

0

No output

** 1

Manual, Fixed format

2

Auto, Fixed format

3

Manual, Format 1 (f34)

4

Auto, Format 1 (f34)

5

Manual, Format 2 (f35)

f41
Accumulated data
output at
accumulated data
display

6

Auto, Format 2 (f35)

** 0

No delay

1

0.5 second

2

1.0 second

3

1.5 seconds

f42
Delay for continual data

4

2.0 seconds

** 0

Not use

f43
Command address

1

Use (Address determined f06)

** 0

Approx. 1 second

f44
Time out

1

No limitation

** 0

CR, LF (0Dh, 0Ah)

f45
Terminator

1

CR (0Dh)

*:

Initial

settings.

6.2. F-Functions

Advertising