5.5.

5.5.

5.5.

5.5. Gravity

C

Gravity C

Gravity C

Gravity Compensation

ompensation

ompensation

ompensation F

F

F

Function

unction

unction

unction

If the scale is used at the calibration location, it is not necessary to perform this function.

If there is a difference of gravity acceleration between the installed location and

calibration location it may cause a weighing error. This function specifies the

gravity acceleration and corrects the span error.

Note

The decimal point is not displayed in the function. Example: 9798 = 9.798 m/s

2

When span calibration is executed, the gravity acceleration correction will

be cleared and the two gravity acceleration values will return to the factory

settings.

Step 1 At the Cal 0 display, press the

MODE

key until g 5et is displayed and press the

ENTER

key to enter the gravity compensation function.

If you want to cancel the current procedure, press and hold the

ON/OFF

key and press

the

ESC

key. Then, no parameters are changed and the calibration mode is finished.

Press the

ON/OFF

key to turn the display off after displaying CanCel.

Step 2 The parameter is displayed with triangle 1. Enter the gravity acceleration of the

calibration location using the

0

-

9

keys. The parameter xxxx is the gravity

acceleration.

ENTER

key...The key to store the new gravity acceleration and proceed to step 3.

ESC

key .......The key to return to g 5et without changing the value.

Step 3 The parameter is displayed with triangle 2. Enter the gravity acceleration of the

installed location using the

0

-

9

keys. The parameter xxxx is the gravity acceleration.

ENTER

key...The key to store the new gravity acceleration and proceed to step 4.

ESC

key .......The key to return to step 2 without changing the value.

Step 4 Now

g xxxx

is displayed. Press the

CAL

key to store the parameters. The Caloff is

displayed. Proceed to step5.
NOTE: When displaying Caloff , press the +/- key while pressing the

ON/OFF

key,

instead of

CAL

key.

Step 5 Press the

ON/OFF

key to turn the display off.

5.5. Gravity Compensation Function