A&D Weighing AD-4407 User Manual

Page 15

Advertising
background image

AD-4407 Weighing Indicator

Page

13

Single Range

Select resolution and decimal point position and format.

Specify the weighing capacity.

Dual Range

<First range> Select the resolution, decimal point position and format.

<First range> Select the weighing range

<Second range> Select the resolution

<Second Range> Specify the weighing capacity

For the range function, refer to “5.3. Weighing Range Function”.

Specifying the Range and Unit

Specifying the Range and Unit

Specifying the Range and Unit

Specifying the Range and Unit

Step 1

The range and unit of measure are displayed.

Range display : 5ingl : single range

dUal : dual range

To change the range function, use the

CLEAR

key

Unit display:

The active unit is displayed. The unit can be changed such as

kg or lb. The calibration is performed with displayed unit.

Press

the

MODE

key to select a unit and to select a unit for calibration,

use the +/- key. The unit for calibration (first unit) is displayed and the

alternate unit (second unit) is blinking.

ENTER

The key to store the data displayed and proceed to the next step.

ESC

The key to proceed to the next step without changing them.

Specifying the Resolution, Decimal Point Position an

Specifying the Resolution, Decimal Point Position an

Specifying the Resolution, Decimal Point Position an

Specifying the Resolution, Decimal Point Position and Format

d Format

d Format

d Format

Step 2 The resolution will be displayed as d 0.1 with decimal point. The indicator displays

triangle

1 and the first unit selected at the previous step. Specify the position of

the decimal point with the

MODE

key. Specify the display format (point or comma)

with the +/- key. Specify the resolution with

CLEAR

key. The decimal point format

is specified in step. The decimal point format for serial data output is selected

using the F-function settings. Press the

ENTER

key to store them and proceed to

the next step. When pressing the

ESC

key, regardless of what is displayed, the

indicator will proceed to the next step without changing them.

5.2. Calibration Procedure