requires a Station Identification (ID code) so that it can clearly be

identified.
1. Select the menu item REMOTE ID in Menu 2.
2. Enter an ID with the UP or DOWN Scroll Keys

(possible values 0 – 999).

3. Proceed to the next menu item with the Radio Button ┌ 1 ┐ (back)

or ┌ 2 ┐ (forward) or exit the menu with Radio button ┌ 3 ┐ (EXIT).

Note

Depress Radio Button ┌ 3 ┐ to exit the menu item without saving any

changes (EXIT).

Factory Control Check ( FCC )

This function checks temperature accuracy of the Rework Station

and allows modifications if necessary. To perform the “FCC” function,

the soldering tip temperature must be measured using an external

temperature measuring instrument, (WA2000) and appropriate type

“K” temperature measuring tip. ( See Accessories List on page 18 )

Control Check at 212 °F / 100 °C

1. Connect the type “K” thermocouple, into the external temperature

measuring instrument, (WA2000).

2. Select the menu item FCC in Menu 2.
3. Depress the DOWN Scroll Key. Control point 212 °F / 100 °C is

selected.

The soldering tip is now heated to 212 °F / 100 °C.

The Heater Control Indicator flashes (a) (21) as soon as the

temperature is constant.

4. Compare the temperatures indicated by the meter with the

indications in the display.

5. Use the UP or DOWN Scroll Key to set the difference between the

value indicated on the external meter and the value indicated on

the Rework Station, (shown in small script in lower part of the

display). Maximum temperature adjustment: ± 72 °F (± 40 °C).

Example:
Display 212 °F, external meter 210 °F: setting S 2
Display 212 °F, external meter 214 °F: setting T 2

Note

Depress Radio Button ┌ 3 ┐ to exit the menu item without saving

changes (EXIT).

6. Depress Radio Button ┌ 2 ┐ (Set) to confirm the value.

The temperature deviation is now reset to 0. Factory Control

Check at 212 °F / 100 °C is now complete.

7. Exit Menu 2 by depressing Radio Button ┌ 3 ┐ twice.

Control Check at 842 °F / 450 °C

1. Connect the type “K” thermocouple, into the external temperature

measuring instrument, (WA2000).