High altitude locations, Power source, Installation (continued) – Shellab BACTRON600 User Manual

Page 12

Advertising
background image

12 |

P a g e

INSTALLATION (CONTINUED)

H

IGH

A

LTITUDE

L

OCATIONS

BACTRON anaerobic workstations are equipped with two (2) digital vacuum controller - display
units that control the atmospheric cycling operations of the pass box and the vacuum pump line
pressure. These devices are calibrated at the factory at an altitude of 179 feet (54.5m) above sea
level, and read gauge pressure rather than absolute pressure. Altitudes around 6500 feet (2000m)
or higher may affect the operation of the vacuum system, depending on ambient conditions. If the
vacuum pump runs continuously or near continuously when installed at a high altitude location,
contact

Sheldon Technical Support

(see page 6) for assistance in adjusting the vacuum gauge

settings.

P

OWER

S

OURCE

Note: The electrical supply to the BACTRON

must conform to all national and local electrical codes.

Always position the workstation so that the users have access to the power cord and can quickly
unplug it in the event of an emergency.

When choosing a location for the BACTRON check that the voltage and ampere requirements on

the workstation’s data plate match those of your wall source

. The source must be an earth

grounded outlet The supplied voltage must not vary more than 10% from the data plate rating.
Damage to the workstation may result if supplied voltage varies more than 10%.
Each
BACTRON workstation is provided with a 110

120VAC 8ft (2.5m) 5-15 NEMA power cord.

These incubators are intended for a 50/60 Hz, 110 - 120 volt applications at the following
amperages:

Model

Amperage

BACTRON300 (BACTRONII)

9 Amps

BACTRON600 (BACTRONIV)

11 Amps

BACTRON900 (BACTRONIV-900)

14 Amps

Use a separate circuit to prevent loss of the unit due to overloading or circuit failure.