OPERATION (CONTINUED)

The Over Temperature Limit System

The OTL is a mechanical heating control backup system included with each incubator, and which
operates independently of the

incubator’s digital temperature c

ontroller board. The OTL is intended

to be set to approximately 1

°

C above the user-selected temperature set point stored in digital

controller board. The OTL prevents runaway heating in the event that the main controller fails while
in its heating mode.

If the incubator’s temperature exceeds to the OTL limit setting, a red pilot light indicator will
illuminate and an audible alarm will buzz. The OTL System will depower the incubator’s heating

element, and continue to do so as long as the temperature remains higher than the OTL setting.

Manometer Pressure Gauge and Check Valve

The water-filled manometer in the workspace chamber serves as a visual pressure gauge, as well
as a venting check valve during instances of excess overpressures.

Figure 7: Manometer Components

An increase in atmospheric pressure inside the workstation drives down water within the
Manometer relative to a pair of reference and refill rings. Under normal operating conditions the

workstation’s overpressure displaces a

-half inch of water (0.5cm). Excessive pressure will cause

the water to bubble as chamber atmosphere is vented through the manometer and out of the
workstation. This helps prevents damage to workstation gaskets. The manometer exhaust vent is a
port consisting of a transparent tube and black O-ring located on the back, right side of the
BACTRON.