Cleaning, User maintenance (continued), Leaning – Shellab BACTRON600 User Manual

Page 56

Advertising
background image

56 |

P a g e

USER MAINTENANCE (CONTINUED)

Warning

: Disconnect the power cord from the wall power supply prior to cleaning or disinfecting the

BACTRON.

Avertissement

: Avant d'effectuer toute entretien de cet appareil, débrancher le cordon secteur de la

source d'alimentation.

If a hazardous material/substance has spilled in the BACTRON

, immediately initiate your site’s

Hazardous Material Spill Containment protocol. Contact your local Site Safety Officer and follow
instructions per the site policy and procedures.

Periodic cleaning and the disinfection are required to prevent microbiological contamination.

Warning

: Never clean or disinfect the BACTRON with alcohol or flammable cleaners.

Avertissement:

Ne jamais nettoyer l'appareil à l'alcool ou avec des nettoyants inflammables.

Note: Do not use spray cleaners or disinfectants that might leak through openings and cracks and

get on electrical components, or that contain solvents that will harm coatings. Do not use
chlorine-based bleaches or abrasives; these will damage the interior surfaces.

C

LEANING

Note: The

BACTRON’s

chamber and incubator should be cleaned and disinfected prior to first use.

1. Remove all items (shelves, racks, and any additional items) from the incubator(s).

2. Clean the workstation

’s interior

with a mild soap and water solution, including all corners.

3. Clean all removable accessories including arm port doors, gaskets, and sleeve assemblies

with a mild soap and water solution.

4. Rinse with distilled water and wipe dry with a soft cloth. Do not use deionized water. See

the

Distilled and Deionized Water

paragraph in the Operation section.

5. Take special care when cleaning around the pass box door alarm sensors and chamber

power outlets to prevent damage.

6. Do not use chloride-based cleaners except Zephiran benzalkonium chloride solution.

Other types may have adverse effects on microbiological samples.

7. Wipe down the interior surfaces with Zephiran. Do not wipe up, allow the Zephiran to

evaporate.

Advertising
This manual is related to the following products: