Single gas connection, Dual gas connection, E 30 – Shellab BACTRON600 User Manual
Page 30: Operation (continued)
30 |
P a g e
OPERATION (CONTINUED)
S
INGLE
G
AS
C
ONNECTION
Follow these steps to connect a single cylinder of anaerobic mixed gas.
1. Locate the T-shaped brass fitting included with the BACTRO
N’
s AMG gas regulator
(normally shipped in the regulator box).
AMG
GAS 1
IN
GAS 2
IN
Figure 14: T-Fitting AMG Figure 15: Single Gas Connections
2. Using the attached short lengths of black OD tubing, connect the fitting to the GAS 1 IN
and GAS 2 IN ports on the top left side of the BACTRON. See figure 13 on the previous
page.
3. Connect a barb adaptor on the T-fitting to the AMG gas regulator using the length of black
OD included with the regulator.
Note: Failure to connect the T-fitting to both the GAS 1 and 2 ports will prevent the pass box auto
cycle from cycling the pass box.
D
UAL
G
AS
C
ONNECTION
To connect AMG and an inert recharging gas such as, nitrogen
(N
2
), complete the following steps:
1. Connect the AMG
regulator’s
tubing to the
BACTRON’s
GAS 1 IN port.
2. Connect the inert recharging gas cylinder regulator line
to the BACTRON’s
GAS 2 IN port.
Note: When plumbed in this configuration, the BACTRON will draw twice from the GAS 2 IN port,
and once from the GAS 1 IN port during pass box auto cycles. This is intended to reduce
pass box AMG consumption. The BACTRON will not draw from the GAS 2 IN port during
manual gas injections, sleeve assembly purges, or other operations.
Figure 16: Dual Gas Connections