23 |

P a g e

CONTROL PANEL OVERVIEW (CONTINUED)

Workspace Incubator Control Panel (BACTRON900 / BACTRONIV-900 Only)

Figure 5: BACTRON900 Workspace Control Panel

The workspace incubator control panel on the BACTRON900 controls the operations of the
workspace

chamber incubator and the incubator’s Over Temperature Limit system.

Main Temperature Control and Green Digital Display

The BACTRON900’s

workspace SET TEMPERATURE control panel comes with a green digital

display that shows the temperature of the incubator accurate to within + 0.1

C. The control consists

of up and down arrow keys for adjusting set point temperatures or placing the display in its
calibration mode.

Heating Activated Light

The clear pilot light located beneath the label HEATING ACTIVATED illuminates whenever the

heating elements are powered and warming the BACTRON900’s

workspace chamber Incubator

Set Over Temperature

This graduated dial sets the temperature set point for the Over Temperature Limit backup system in
the workspace chamber of the BACTRON900. The OTL System operates independently of the
Main Temperature Controller, and prevents uncontrolled heating of the workspace incubator in the
event of a temperature control board failure. For more details, please see the explanation of

the

Over Temperature Limit System

in the Operation Section.

OTL Light

This pilot light is marked OVERTEMPERATURE ACTIVATED and is located directly below the
Over Temperature Limit control. The light will turn on and a buzzer alarm will activate when the
OTL System has taken control of the incubator. Under normal operating conditions this light should
not illuminate.