Nput connections – ADLINK PXI-2208 User Manual

Page 42

Advertising
background image

30

Signal Connections

Referenced Single-ended (RSE) Mode

In referenced single-ended mode, all input signals are con-
nected to the ground provided by the DAQ-/DAQe-/PXI-2204/
2205/2206/2208 card. This is suitable for connections with
floating signal sources. Figure 3-1 shows an illustration. Note
that when more than two floating sources are connected, these
sources will be referenced to the same common ground.

Figure 3-1: Floating Source and RSE Input Connections

Non-Referenced Single-ended (NRSE) Mode

To measure ground-referenced signal sources, which are con-
nected to the same ground point, you can connect the signals
in NRSE mode. Figure 3-2 illustrates the connection. The sig-
nals local ground reference is connected to the negative input
of the instrumentation Amplifier (AISENSE pin on CN1 connec-
tor), and the common-mode ground potential between signal
ground and the ground on board will be rejected by the instru-
mentation amplifier.

Figure 3-2: Ground-referenced Sources and NRSE Input Connections

AIn

AIGND

n = 0, ...,63

Floating
Signal
Source

Instrumentation
Amplifier

To A/D
Converter

Input Multipexer

V1

V2

+

-

+

-

CN1

AIn

AISENSE

n = 0, ...,63

Ground-
Referenced
Signal Source

Instrumentation
Amplifier

To A/D
Converter

Input Multipexer

V1

V2

+

-

+

-

V

cm

Common-
mode noise &
Ground
potential

Advertising
This manual is related to the following products: