72

Operation Theory

Figure 4-31: Analog Trigger Block Diagram

The trigger signal is generated when the analog trigger condition
is satisfied. There are five analog trigger conditions in the DAQ-/
DAQe-/PXI-2204/2205/2206/2208 card. The DAQ-/DAQe-/PXI-
2204/2205/2206/2208 card uses two threshold voltages,
Low_Threshold and High_Threshold to build the five different trig-
ger conditions. You can configure the trigger conditions easily by
software.

Trigger level digital setting Trigger voltage

0xFF

9.92V

0xFE

9.84V

0x81

0.08V

0x80

0

0x7F

-0.08V

0x01

-9.92V

Table 4-7: Analog Trigger SRC1 (EXTATRIG) Ideal Transfer Characteristic