Figure 3-9: Middle Trigger Mode Acquisition

3.4.5

Acquisition with Re-Triggering

A digitizer acquires a trace of N samples/channel for a single
acquisition. Re-Trigger mode can also be set to automatically
acquire R traces, containing N*R samples/channel of data, without
additional software intervention.

The Re-Trigger setting can be used for Post-Trigger and Delayed-
Trigger modes, with different limitations on the spacing between
trigger events in each mode. Trigger events arriving too close to
the previous instance will be ignored by the digitizer.

X

In Post-Trigger mode, the minimum spacing between trigger
events is N+1

X

In Delayed-Trigger mode, the minimum spacing between
trigger events is (N+D)+1, where D is the number of the
delayed setting

Figure 3-10: Re-Trigger Mode Acquisition

Time

Operation start
Acquisition start

Trigger

Data

Acquisition stop
Data transfer to system begins

N samples

M samples

Trigger event occurs