Sense[1..n]:correction:offset [:magnitude – EXFO IQS-1500 Power Meters for IQS-500/600 User Manual
Page 140

SCPI Command Reference
132
IQS-1100/IQS-1500
Product-Specific Commands—Description
:SENSe[1..n]:CORRection:OFFSet
[:MAGNitude]
Description
This command sets an offset value. The units are
w/w by dÈfault.
If no channel was specified, the default channel
used is 1.
Syntax
:SENSe[1..n]:CORRection:OFFSet[:MAGNitude]
<wsp><CorrectionOffset[<wsp>W/W|DB]>|
MAXimum|MINimum|DEFault
Parameter(s)
CorrectionOffset:
The program data syntax for <CorrectionOffset>
is defined as a <numeric_value> element
followed by an optional <SUFFIX PROGRAM
DATA> element. The allowed <SUFFIX
PROGRAM DATA> elements are: W/W|DB. The
<CorrectionOffset> special forms MINimum,
MAXimum and DEFault are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
DEFault allows the instrument to select a value
for the <CorrectionOffset> parameter.
Sets the offset for the specified channel.
Example(s)
SENS:CORR:OFFS:MAGN 2.0
See Also
SENSe:CORRection:OFFSet:[MAGNitude]?