Calibrating Other Power Meters with the IQS-1500
70
IQS-1100/IQS-1500
Entering Calibration Uncertainties
The remaining uncertainties are optional, but to have the best possible
estimation, all uncertainties must be considered.
To set the uncertainties values for your calibration test:
1. Select the wavelength for which you want to enter uncertainties.
2. Enter the Type A uncertainty contribution in the box if needed.
3. Select the Other check box so that the entry is included in the
calculations.
4. Enter the Type B uncertainty values according to the calculations
performed earlier.
5. Enter the Reference Meter expanded uncertainty in the corresponding
box.
Once you are done entering the uncertainty values, click Next to proceed
to the next step.