Calibrating Other Power Meters with the IQS-1500

70

IQS-1100/IQS-1500

Entering Calibration Uncertainties

The remaining uncertainties are optional, but to have the best possible
estimation, all uncertainties must be considered.

To set the uncertainties values for your calibration test:
1.
Select the wavelength for which you want to enter uncertainties.

2. Enter the Type A uncertainty contribution in the box if needed.

3. Select the Other check box so that the entry is included in the

calculations.

4. Enter the Type B uncertainty values according to the calculations

performed earlier.

5. Enter the Reference Meter expanded uncertainty in the corresponding

box.

Once you are done entering the uncertainty values, click Next to proceed
to the next step.