Maintenance
Power Meters
91
Cleaning Fixed Connectors
7. Slowly insert the cleaning tip into the connector until it reaches the
ferrule inside (a slow clockwise rotating movement may help).
8. Gently turn the cleaning tip one full turn.
9. Continue to turn as you withdraw the cleaning tip.
10. Repeat steps 7 to 9, but this time with a dry cleaning tip (2.5 mm tip
provided by EXFO).
Note: Make sure you don’t touch the soft end of the cleaning tip and verify the
cleanliness of the cotton tip.
11. Throw out the cleaning tips after one use.
I
MPORTANT
Alcohol may leave traces if used abundantly. Avoid contact between 
the tip of the bottle and the cleaning tip, and do not use bottles 
that distribute too much alcohol at a time.