Hanna Instruments HI 83749 User Manual

Page 11

Advertising
background image

11

REMOVING AIR BUBBLES

Air bubbles present in the sample will cause erroneous high turbidity readings.

• To remove air bubbles fill the cuvet with wine sample and close the cap tightly. Shake the

cuvet gently to create pressure. Allow the cuvet to stand for a few minutes and gently invert it

several times. Check that no air bubbles are visible, otherwise shake again and repeat the

procedure above.

• Alternatively use an ultrasonic bath to degas the wine sample.

• In order to measure the exact volume of bentonite suspension with the 1 mL syringe, push the

plunger completely into the syringe and insert the tip into the solution. Pull the plunger up to

above the 0.0 mL mark. Take out the syringe and clean the outside of the syringe tip. Then,

adjust the plunger to the 0.0 mL mark (the lower edge of the seal must be exactly on the 0.0 mL

mark). Be sure that no drops are hanging on the tip of the syringe, if so eliminate them.

Then, to add exactly 0.25 mL of bentonite suspension, keep the syringe in vertical position over

the cylinder and push the plunger down until the lower edge of the seal is exactly on the 0.25 mL

mark. Now the exact amount of 0.25 mL has been added to the cylinder, even if the tip still

contains some solution.

Repeat the entire procedure to measure 0.50 mL, 0.75 mL and 1 mL of bentonite suspension.

MEASUREMENT TIPS

• For a correct filling of the cuvet: the liquid in the cuvet forms a

convexity on the top; the bottom of this convexity must be at the

same level of the 10 mL mark.

• For dosing the Bentocheck reagent, we recommend to use the

supplied Hanna automatic pipette (HI 731341 - 1000 µL).

For a correct use of the Hanna automatic pipette, please follow

the related Instruction Sheet.

0.25 mL

0.50 mL

or

Hanna

automatic pipette

30

alternated with

• Add HI 83749-0 Bentocheck reagent, mix and

wait for 1 minute. Then take a new reading (T2).

• If the difference between T1 and T2 is less than

10%, the instrument gives a beep of 1 second

and displays alternating the turbidity value

(NTU) and the difference (%).

• If the difference between T1 and T2 is more

than 10%, the instrument just displays T2 in

the primary LCD and T1 in the secondary LCD.

ACTIVATING COMPARATIVE MODE

For the determination of bentonite requirement you

can activate Comparative mode for an automatic

comparison of turbidity values.

• To edit the Comparative mode, press CAL/RCL

when “Activating Comparative mode” panel is

displayed.

The parameter setting and “CFM” tag will start

blinking.

• Press the UP or DOWN keys to set the

Comparative mode on or off.

• Press LOG/CFM to save the setting. The new

selected option will be displayed on the LCD.

Alternatively, press CAL/RCL to exit without

saving the new settings.

Note: when Comparative mode is set on, in the

Measurement mode, the secondary display will

show the reference value

or

Advertising
This manual is related to the following products: