Hanna Instruments HI 83749 User Manual

Page 15

Advertising
background image

15

DELETE CALIBRATION

HI 83749 is factory calibrated. It is possible to

restore factory calibration by deleting last performed

calibration.

To delete last calibration, follow the next steps:

• Enter the GLP feature by pressing GLP .

The date of the last calibration will be displayed

on the LCD (e.g. 2005.02.23).

• Press READ to scroll through the information

related to calibration. The last panel is the one

with “Delete Calibration”.

• Press SETUP/DEL to delete the current

calibration. After deletion the instrument will

automatically return to measurement mode and

the factory calibration is restored.

26

• Measurement date in YYYY.MM.DD format.

• Measurement time in hh:mm format.

• Delete the last record panel (only for last record).

• Delete all records.

DELETE LAST RECORD
To delete the last record, scroll through the log menu

until the delete last record panel is displayed.

• To delete the last record, press SETUP/DEL when

the “Delete last records” panel is displayed.

• The instrument asks for confirmation. Press the

LOG/CFM key to confirm the last record is

deleted. To abort the delete function, press

READ instead of LOG/CFM.

• After the record is deleted, the instrument goes

immediately to the first panel of the previous

record. If the log becomes empty, “----” will be

displayed for one second and the instrument

will return to measurement mode.

DELETE ALL RECORDS
To delete all records, scroll through the log until

delete all records panel is displayed.
• To delete all records press SETUP/DEL when the

“Delete all records” panel is displayed.