DISPLAY DESCRIPTION

1) Battery icon. This icon appears when the status of the battery is displayed or when the battery

voltage is getting low.

2) Hour glass icon. It is displayed when the instrument performs an internal checkup.

3) Lamp and read status indicator.

4) Four digit main display.

5) NTU measurement units. When average or continuous mode is selected, the “NTU” tag blinks

for each new displayed value. For conversions in other units see Measurement Units section.

6) AVG icon appears when the Signal Average Mode is selected.

7) Four digit secondary display.

BEEPER

A long beep indicates an error or an invalid key pressed. A short beep means that the current

operation is confirmed.

11) CAL/RCL: this is a bi-functional key. Just press to enter/exit calibration or in setup mode to

start/stop editing a parameter. Hold the key for 3 seconds to enter/exit viewing log content.

12) LOG/CFM: press to save a record or to confirm the selected option.

13) SETUP/DEL: press to enter/exit setup. The DEL function is available in Log Recall mode to

delete one or all records. In GLP it is used to restore factory calibration.

14) READ : press to start a measurement. Hold the key to make a continuous measurement. In

Log Recall mode it is used to view the content of a record. In GLP it is used to view all

available information. In Setup, during date or time editing, it is used to select day, month or

year and hour/minutes.

CONNECTORS DESCRIPTION

15) AC adapter connector.

16) RS232 connector, to be used with serial

cable to trasfer data to PC.

17) Tag reader.

18) USB connector.

32

or

or

or

SHOW / HIDE THE TIME

You can choose between showing or hiding the

current hour and minutes on the secondary LCD.

• To set hiding or showing the time, press

CAL/RCL when the “Show/hide time” panel is

displayed.

The time show status and the CFM tag will start

blinking.

• Press the UP or DOWN keys to set Lcd / hide for

time.

• Press LOG/CFM to save the change. The new

selected option will be displayed on the LCD.

Alternatively, press CAL/RCL to exit without

saving the changes.

SET THE DATE

The HI 83749 turbidimeter has a built-in real time

clock (RTC). The RTC time is used to generate a

unique time stamp for each recorded value and to

automatically store the last calibration date. The

current time can be displayed on the LCD when the

instrument is in measurement mode.

• To set the current date, press CAL/RCL when the

“Set date” panel is displayed. The date format

is YYYY.MM.DD. The last two digits of the year

value and CFM tag will start blinking.

• Press the UP or DOWN keys to set the year value.

• Press LOG/CFM or READ to start editing the

month value. The month value will start

blinking.

• Press the UP or DOWN keys to set the month

value.