Hanna Instruments HI 83749 User Manual

Page 17

Advertising
background image

17

AVG (SIGNAL AVERAGE MODE)

It is recommanded to select this measurement mode when you work with unstable samples that

contain suspended particles of different dimensions.

In the AVG mode the instrument takes 20 measurements in a short period of time and displays the

updated averaged value.

To select the averaged measurement mode press

the AVG key and the AVG icon will appear on

the LCD.

Note: to return to the normal measurement mode

just press again the AVG key.

• Press READ and “----” will blink on the dis-

play. After a few seconds the instrument displays

the first reading. The meter continues updating

the readings until the “Lamp and Read Status”

indicator turns off. The final displayed value is

the averaged reading of turbidity in NTU.

CONTINUOUS MEASUREMENT

This measurement mode can be used to verify how fast suspended parts settle out. Please verify first

that the AVG mode is turned off (see instructions above). To make continuous measurement keep the

READ key pressed until the desired number of measurements are taken.

The last value remains on the display after the READ key is released.

24

LOGGING

The log function is active after a valid measurement

is obtained (no errors).
• To log a value, press LOG/CFM when the

measurement result is displayed.

The instrument asks to READ TAG for identification

of the sampling location. The location for the new

record is also displayed on the secondary LCD.

• To read the ID code for the sampling location

identification, simply touch the iButton

®

tag with

the matching connector, located on the back of the

instrument (see Connectors Description, page 9).

Alternatively, press again LOG/CFM to store the

record without the tag ID code.

• If the tag is successfully read, the instrument will

beep once, displaying the unique hexadecimal

code of the tag, and store the data.
After data is stored, the instrument returns to

measurement mode.

Notes: • If the tag is not read within 20 seconds,

the logging procedure is canceled.

• A measurement can be stored only once.

Also an over range value can be stored.

• If less than ten free records are available, the

“LOG” tag will blink while storing data.

• If the log memory is full, the “LoG FULL” message

will appear for a few seconds on the LCD and the

instrument will return to measurement mode

without storing the new record.

To store a new record, delete one or more records.

HI 83749 has a log space for up to 200 records. With each measurement, the date, time and tag ID

is stored. In this way, each record is fully characterized and can be easily analyzed when downloading

data on the PC application (HI 92000).

or

or

LOGGING

Advertising
This manual is related to the following products: