Hanna Instruments HI 83749 User Manual

Page 14

Advertising
background image

14

• Remove the 100 NTU standard cuvet from the

meter and insert the 500 NTU standard cuvet

into the holder.

• Close the lid and press READ . “----” and the

“Lamp and Read Status” indicator will blink

again while making the reading.

• At the end of the measurement, the four-point

calibration is completed and the instrument

returns automatically to measurement mode.

OUT CAL RANGE FUNCTION

The instrument has an Out Cal Range function to

alert the user (with “cal” blinking message) when

a measurement is made outside of the calibration

range.

CALIBRATION ERRORS

• If the read value during calibration is too far

from the set value, the instrument will show

“-LO-” or “-HI-” error.

• If the calculated calibration coefficients are off

specifications, the “CAL Err” message is

displayed.

or

27

• Press GLP to enter or exit GLP data consulting.

Several functions are available when in GLP

menu.

Press READ to scroll through the GLP data.

The following GLP panels can be viewed.

• The last calibration date, in YYYY.MM.DD

format. If no calibration was performed, the

factory calibration message, “F.CAL”, will be

displayed on the LCD.

• The time of the last calibration in hh:mm format.

• First calibration point: 0.00 NTU if skipped or

the actual read value (e.g. 0.01 NTU).

• Second calibration point.

• The instrument asks for confirmation. Press the

LOG/CFM key to confirm all records are deleted.

To abort the delete function, press READ

instead of LOG/CFM.

• After all records are deleted the instrument

returns to measurement mode.

The GLP feature allows the user to view last calibration data. Also the user calibration can be deleted.

or

or

GOOD LABORATORY PRACTICE (GLP)