Hanna Instruments HI 83749 User Manual

Page 7

Advertising
background image

7

Range

0.00 to 9.99 NTU

10.0 to 99.9 NTU

100 to 1200 NTU

Range Selection

Automatically

Resolution

0.01 NTU from 0.00 to 9.99 NTU

0.1 NTU from 10.0 to 99.9 NTU

1 NTU from 100 to 1200 NTU

Accuracy

±2% of reading plus 0.05 NTU

Repeatibility

±1% of reading or 0.02 NTU, whichever is greater

Stray Light

< 0.05 NTU

Light Source

Tungsten filament lamp

Light Detector

Silicon Photocell

Method

Ratio Nephelometric Method.

Display

60 x 90mm backlit LCD

Calibration

Two, three or four points calibration

LOG Memory

200 records

Serial Interface

RS232 or USB 1.1

Environment

0 to 50°C (32 to 122°F); max 95% RH non-condensing

Power supply

4 x 1.5V AA alkaline batteries or AC adapter

Auto Shut-off

After 15 minutes of non-use

Dimensions

224 x 87 x 77 mm (8.8 x 3.4 x 3.0”)

Weight

512 g (18 oz.)

SPECIFICATIONS

34

SET INSTRUMENT ID

The instrument ID is a four digit number that can

be edited by the user. The instrument ID is

downloaded on the PC application, together with

the logged data. By setting a different ID for each

instrument it is possible to mix information from

many turbidimeters into the same database.

• To set the instrument ID, press CAL/RCL when

the “Set instrument ID” panel is displayed.

The default instrument ID is 0000. The existing

ID value and the CFM tag will start blinking.

• Press the UP or DOWN keys to set the new

instrument ID. By pressing and holding the UP

or DOWN keys, the changing speed wil increase.

• Press LOG/CFM to save the change. The new

instrument ID will be displayed.

Alternatively, press CAL/RCL to exit without

saving the changes.

SET BAUD RATE

The HI 83749 has a RS232 and a USB link. When

the USB connection is used, the RS232 connection

becomes inactive.

To successfully communicate with the PC, the same

baud rate must be selected on the instrument and

on the PC application. The available baud rates are

1200, 2400, 4800 and 9600.
• To set the baud rate, press CAL/RCL when the

“Set baud rate” panel is displayed.

The parameter value and the CFM tag will start

blinking.

• Press the UP or DOWN keys to select the new

baud rate value.

• Press LOG/CFM to save the change. The new

selected baud rate will be displayed.

Alternatively, press CAL/RCL to exit without

saving the changes.