Hanna Instruments HI 83749 User Manual

Page 13

Advertising
background image

13

• Remove the <0,10 NTU standard cuvet and

place the 15.0 NTU standard cuvet into the

holder.

• Close the lid and press READ . “----” and the

“Lamp and Read Status” indicator will blink

again while making the reading.

• At the end of the reading, the third calibration

point (100 NTU) and “CAL P.3” is displayed. If

desired, the value can be changed using the

UP or DOWN keys.

Note: At this moment it is possible to exit

calibration by pressing CAL/RCL. The instrument

will memorize the two-point (<0.10 and 15.0

NTU) calibration data and will return to

measurement mode.

• Remove the 15.0 NTU standard cuvet from the

meter and place the 100 NTU standard cuvet

into the holder.

• Close the lid and press READ . “----” and the

“Lamp and Read Status” indicator will blink

again while making the reading.

• At the end of the reading, the fourth calibration

point (500 NTU) and “CAL P.4” is displayed. If

desired, the value can be changed using the

UP or DOWN keys.

Note: At this moment it is possible to exit

calibration by pressing CAL/RCL. The instrument

will memorize the three-point (<0.10, 15.0

and 100 NTU) calibration data and will return

to measurement mode.

28

• Third calibration point (if available).

• Fourth calibration point (if available).

• Delete calibration panel.

To delete calibration:

• Press SETUP/DEL when the instruments displays

the “Delete calibration” panel.

The user calibration will be deleted and the factory

calibration will be restored. The instrument will

enter automatically in idle mode.

Advertising
This manual is related to the following products: