12

It is recommended to calibrate the meter only with Hanna ready-to-use calibration standard solutions.

Alternatively, formazin standards can be used. The prepared formazin solutions should be close

to the default calibration points. The first point must be near 0 NTU, the second point between

10 and 20 NTU, the third point between 50 and 150 NTU and the fourth point between 400

and 600 NTU.

CALIBRATION

Calibration can be performed in two, three or four points. It is possible to interrupt calibration procedure

at any time by pressing ON/OFF.

CALIBRATION PROCEDURE

• Turn the instrument on by pressing ON/OFF.

When the LCD displays “----”, the instrument

is ready.

• Enter calibration mode by pressing CAL/RCL.

The display will show “CAL P.1”.

• Place the <0.10 NTU standard cuvet into the

holder.

• Close the lid and press READ . “----” and the

“Lamp and Read Status” indicator will blink

on the display.

Alternatively, press LOG/CFM to skip the first

calibration point.

• Then the LCD will show the second calibration

point (15.0 NTU) and “CAL P.2”, while “READ”

is blinking.

Note: If you’re using different calibration

standards, change the displayed value by

pressing UP or DOWN keys until the display

shows the desired value.

or

29

The Setup mode allows the user to view and modify the instrument parameters.

The blinking “CAL” tag appears during setup mode suggesting to press CAL for editing parameters.

SETUP

SET REFERENCE VALUE

During determination of bentonite requirement for

stabilization of the wine, it can be useful to set a

reference value and let the instrument automatically

compare turbidity values.

• When the display shows the first reading (T1),

you can store the result as Reference Value. Press

SETUP/DEL to enter the “Set Reference Value”

panel. “CAL” will blink and “SET rEF” will be

displayed.

• To set the reference value, press CAL/RCL. Then

the value and “CFM” tag will start blinking.

• At this point the displayed reference value can

be modified using the UP and DOWN arrow keys.

• Press LOG/CFM to store the value.

Note: To exit without saving the reference value,

press SETUP/DEL and the meter will return to

the measurement panel.

• To activate the Comparative mode, press the

AVG key and set on the Comparative mode

(see page 30).

• Press SETUP/DEL to return to Measurement

mode. The display will show the reference value.

or

or

• To enter/exit SETUP, press SETUP/DEL.

• To select the parameter to be edit, press UP or

DOWN keys until the desired panel is displayed.