Chapter 3: Getting Started


Your Protector Radioisotope Fume Hood has been
tested at the factory per ASHRAE 110-1995. All hoods
achieve an “as manufactured rating” of less than 0.10
part per million (ppm) at 4 liters per minute (lpm); AM
<0.10 (Consult Labconco for individual fume hood
ratings.) For “field use” ASHRAE testing contact the
Labconco Ventilation Ventures Team or Customer
Service for the name of a certified on-site contractor.

NOTE: Face velocity profiles and
smoke testing should be done
periodically to ensure safe
performance
.


Your Next Step


After your fume hood has been installed and certified,
you are ready to proceed to Chapter 4: Performance
Features and Safety Precautions
.












!