Chapter 3: Getting Started
Your Protector Radioisotope Fume Hood has been
tested at the factory per ASHRAE 110-1995. All hoods
achieve an “as manufactured rating” of less than 0.10
part per million (ppm) at 4 liters per minute (lpm); AM
<0.10 (Consult Labconco for individual fume hood
ratings.) For “field use” ASHRAE testing contact the
Labconco Ventilation Ventures Team or Customer
Service for the name of a certified on-site contractor.
NOTE: Face velocity profiles and
smoke testing should be done
periodically to ensure safe
performance.
Your Next Step
After your fume hood has been installed and certified,
you are ready to proceed to Chapter 4: Performance
Features and Safety Precautions.
!