Labconco Protector Stainless Steel Radioisotope Laboratory Fume Hoods 9682200 User Manual

Page 51

Advertising
background image

47

Product Service 1-800-522-7658

Chapter 8: Troubleshooting


PROBLEM

CAUSE

CORRECTIVE ACTION

Vertical-rising

sash

frame is distorted.

Place horizontal sliding glass
symmetrically and pull sash down to air
foil. Straighten damaged frame.

Cable is frayed or has
slipped off the
pulleys.

Re-install, cable must be replaced
immediately if damaged.

Electrical duplex
outlets no longer
have power

Wires not connected
or faulty duplex.

Check wire connection or replace
duplex.

Circuit

breakers

tripped in building
electrical supply.

Reset circuit breakers.

Service valves no
longer operate

Faulty building
supply.

Inspect building supply shut off valves
and appropriate pressures below 40
PSI.

Valve no longer
operates.

Replace valve and check for leaks.

Supply line or outlet
line has leaks.

Inspect line for leaks and fix any
leaking plumbing connections.