Once the site for your Gas Purifier is properly prepared, you are ready to
unpack, inspect, and install your system. Read this chapter to learn how to:

Unpack and move the Gas Purifier.

Set up the Gas Purifier with the proper supporting structure.

Connect the electrical supply.

Connect the plumbing lines for circulation, 4% hydrogen/inert gas

mixture, vacuum, and exhaust vent.

Start-up the Gas Purifier for removal of oxygen and moisture.

Regenerate the Gas Purifier catalysts after 1-3 months of use.


Each Gas Purifier model weighs 75 lbs each (34 kg). The
shipping container allows for lifting with a mechanical lift
truck or floor jack. If you must lift the Gas Purifier
manually, follow safe-lifting guidelines. Do not lift by the
upper sheet metal façade and front panel as damage can
occur.

Unpacking the Gas Purifier

We recommend that you do not remove the Gas Purifier from its shipping
container until it is ready to be placed into its final location.

Carefully remove the shrink-wrap on the Gas Purifier and inspect it for
damage that may have occurred in transit. If damaged, notify the delivery
carrier immediately and retain the entire shipment intact for inspection
by the carrier.

!