Labconco AtmosPure Re-Gen Gas Purifiers 5121845 User Manual

Page 31

Advertising
background image

Chapter 4: Features and Safety Precautions

Product Service 1-800-522-7658

27

11.) The Gas Purifier automatically compensates for power outages and

restarts the regeneration process if interrupted for 10 minutes or
longer. This feature assures complete regeneration of the copper
catalyst and molecular sieve as the system would cool and not
regenerate properly.

12.) Leak tight. Each Gas Purifier has been factory tested with a helium

mass spectrometer and found to have no detectable leaks greater
than 1 x 10

-6

cc/sec.

13.) Each Gas Purifier ships fully regenerated and is ready to use. The

Gas Purifier has the capacity to remove 13,500 ppm oxygen and 18
million ppm moisture.

14.) The Gas Purifier may be programmed by selecting auto mode and

connecting it to an accessory oxygen or moisture monitor. The
back of the Gas Purifier has an Auto Mode Moisture/Oxygen
Connector (9-pin) for connection to the oxygen or moisture
monitor relays. Refer to Labconco accessories for oxygen
monitors, moisture monitors, and the Oxygen/Moisture Monitor
Cable (9-pin).

15.) The Gas Purifier inlet/outlet connections are 3/4" compression

fittings. Connection accessory kits are available with either
neoprene lined flexible hose or hard plumbed stainless steel tubing.
All glove boxes were tested with the neoprene lined hose. An
accessory 3/4" to 3/8" adapter kit is also available for connection
to polyethylene glove boxes and pre-2011 fiberglass/stainless steel
glove boxes.

16.) The Gas Purifier’s compact design easily rests beneath the glove box

on Labconco’s base stands. Dimensions are 14.9"W x 21.9"H x
31.8"D.

17.) The Gas Purifier can be connected to one vacuum pump, shared

with the glove box and powered by the glove box vacuum pump
outlet receptacle. A controlled atmosphere communications port
(C.A. COM 6-pin) on the back of the Gas Purifier allows for
shared access of the vacuum pump when used with Controlled
Atmosphere glove boxes with Auto Pressure Control. A separate
accessory C.A. COM. Cable with 6-pin connectors is available for
communication between the Gas Purifier and the Auto Pressure
Controller of the glove box. The cable must be used to share the
one vacuum pump.

Advertising