Chapter 3: Getting Started

Product Service 1-800-522-7658

22

Step 9 – The regeneration procedure has now been completed. The blower
button needs to be turned on to begin circulating the inert gas in the glove
box through the Gas Purifier.

AUTOMATIC REGENERATION PROGRAMMED FUNCTIONS

FUNCTION

TIME

HEATER

SOLENOID VALVES

4% HYDROGEN

INLET

VENT

VACUUM

H

2

Set Initial

Purge (manual

adjustment of 4%

hydrogen/inert gas

flow rate)

2 MIN.

ON

OPEN

OPEN

CLOSED

HEAT

3 HRS.

ON

CLOSED

OPEN

CLOSED

H

2

PURGE

2 HRS.

ON

OPEN

OPEN

CLOSED

VACUUM

Continues until

manually

switched off by

operator pushing
Start/Stop button

8 hr. (min.)

OFF

CLOSED

CLOSED

OPEN

VACUUM
RELEASE

Momentary (by

operator)

OFF

CLOSED

OPEN

MOMENTARILY

CLOSED


NOTE: The Gas Purifier has optimized programming of each function as
shown in the above chart.

Electrical Power Outage:

The regeneration system automatically compensates for power outages. If
the outage is longer than 10 minutes, the regeneration system will reset and
start over.