Regeneration Control

Regeneration is controlled by a microprocessor which automatically
sequences the solenoid valves during the regeneration cycle. The
regeneration cycle restores the molecular sieve and copper catalyst. It is
programmed for the following time functions:

Initial 2 minute purge to allow adjustment of 4% hydrogen/inert gas

mixture flow rate to 10-25 CFH. (10 CFH Helium/4% Hydrogen, 20
CFH Nitrogen/4% Hydrogen, 25 CFH Argon/4% Hydrogen).

Heat cycle – 3 hours.

Purge cycle – 2 hour – 4% hydrogen/inert gas mixture flows through

purifier (heat also remains on).

Vacuum cycle – 8 hours minimum (continues until manually shut-off

and red LED flashes).


The control panel contains push button switches for automatic cycling (H

2

SET, HEAT, H

2

PURGE, VACUUM). The lighted display shows which

stage the regeneration process is in.

NOTE: Refer to Chapter 3, Chapter 4 and Chapter 5 for more detailed
descriptions of performance features, control functions, operating
instructions, and safety features.