Protector CA (Stainless Steel) Oxygen Removal with AtmosPure Re-Gen Gas Purifier

653.6, 220 cycles stop and

AtmosPure circulation blower

starts and isolation valves are

opened.

1.1

0.8

0.7

210000

42000

10000

3300

55.7

26.2

14.1

7.7

4.2

2.4

1.5

0.1

1

10

100

1000

10000

100000

1000000

0

50

100

150

200

250

Time (Minutes)

O

xy

gen (

PP

M)


This chart shows oxygen level decrease from 210,000 ppm at atmospheric
conditions to 0.7 ppm after 210 minutes. Two separate calibrated oxygen
monitors confirmed results.

*

*

Results were attained at Labconco Corporation under controlled conditions. Actual results

may vary.