Appendix D: Oxygen & Moisture Removal Rates

Product Service 1-800-522-7658

51

Protector CA (Fiberglass) Moisture Removal with AtmosPure Re-Gen Gas Purifier

2.3

600

50.1

3.5

4.6

6.7

10.6

1

10

100

1000

1

10

100

1000

10000

Time (Minutes)

Moist

ur

e (PP

M)


This chart shows moisture level decrease from 600 ppm to 2.3 ppm after
9875 minutes. Two separate calibrated moisture monitors confirmed results.

*

*

Results were attained at Labconco Corporation under controlled conditions. Actual results

may vary.