Labconco AtmosPure Re-Gen Gas Purifiers 5121845 User Manual

Page 23

Advertising
background image

Chapter 3: Getting Started

Product Service 1-800-522-7658

19




Figure 3-7

AtmosPure Re-Gen Control Panel Features

Indicator Lights show each
step of the 13 hour
regeneration process.

Start/Stop Regeneration
button to start and stop the
regeneration process.

Vacuum release button used
momentarily for less than one
second after the regeneration
process is complete.

Main Power
ON/OFF
Switch

Two 3/4"
manually
operated
Isolation Ball
Valves to
isolate Gas
Purifier from
the glove box
during
regeneration.

Flow Control
Valve to set 10-25
CFH flow rate of
4% hydrogen/inert
gas mixture during
regeneration.

Indicator Lights flash
indicating proper position of
the 3/4" manually operated
isolation valves.

Blower button switches Gas
Purifier blower between on, off,
and auto.
ON: Blower on continuously.
OFF: Blower off.
AUTO: Blower controlled by
oxygen or moisture monitors
with high and low range limits
enabled. Accessory Oxygen/
Moisture Monitor Cable (9-pin)
is required.

Advertising