Appendix d: oxygen & moisture removal rates – Labconco AtmosPure Re-Gen Gas Purifiers 5121845 User Manual

Page 51

Advertising
background image

Appendix D: Oxygen & Moisture Removal Rates

Product Service 1-800-522-7658

47

Protector CA (Stainless Steel) Moisture Removal with AtmosPure Re-Gen Gas Purifier

1.0

1.8

2.4

3.5

8.5

15.9

11253, 40% RH

1560

87.5

167.5, 220 cycles stop and

AtmosPure circulation blower

starts and isolation valves are

opened.

445

696

1

10

100

1000

10000

100000

1

10

100

1000

10000

Time (Minutes)

Moist

ur

e (PP

M)


This chart shows moisture level decrease from 11,253 ppm (40% relative
humidity) to 1.0 ppm after 8965 minutes. Two separate calibrated moisture
monitors confirmed results.

*

*

Results were attained at Labconco Corporation under controlled conditions. Actual results