After each use, Cleaning – Laerdal CPRmeter User Manual

Page 18

Advertising
background image

18

19

After each Use

After use on a patient, the CPRmeter may be contaminated and should be
handled appropriately.

1. Place the contaminated CPRmeter in a plastic bag until it can be cleaned.

Do not insert a contaminated CPRmeter into the red sleeve.

2. If it is visibly soiled, wipe the CPRmeter with a soft cloth or paper towel to

remove as much contamination as possible.

3. Remove the CPRmeter patient adhesive from the back of the CPRmeter.

4. Clean the CPRmeter as described under Cleaning.

5. Examine the vent membrane in the center of the yellow rear cover. If the

vent membrane is soiled or damaged, see Replacing the Rear Cover.

6. Check the exterior of the CPRmeter for signs of damage. Contact Laerdal

to arrange for replacement if needed.

7. Apply a new CPRmeter patient adhesive to the CPRmeter as described in

Chapter 2, Setup.

Cleaning

If the CPRmeter has been used in a training situation, it may be wiped using
an alcohol wipe (70% ethanol solution).

If the CPRmeter has been used in a clinical situation, clean it as follows:

1. Clean and scrub the exterior using a mild detergent and either a soft cloth

or toothbrush until the surfaces are visually clean.

2. Wipe the exterior with a soft cloth dampened in lukewarm water.

3. Clean the exterior using a 0.55% solution of ortho-phthalaldehyde. Spray the

solution on to cover all exterior surfaces, and allow to sit for a minimum of
10 minutes. An alternative cleaning agent is isopropyl alcohol (70% solution).

4. Wipe the exterior with a clean soft cloth dipped in water. Allow to dry

completely.

WARNING: The above procedure does not ensure disinfection. Cleaning

the device with 0.55% solution of ortho-phthalaldehyde ensures that the
CPRmeter is thoroughly cleaned.

CAUTION: Do not immerse the CPRmeter in water, hold it under running

water, or allow moisture to penetrate it. Do not sterilize the CPRmeter.

Advertising