Electromagnetic emmissions – Laerdal CPRmeter User Manual

Page 30

Advertising
background image

30

31

30

† The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are

6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz;

and 40,660 MHz to 40,700 MHz.

‡ Field strengths from fixed transmitters, such as base stations for radio (cellular/

cordless) telephones and land mobile radios, amateur radio, AM and FM radio

broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess

the electromagnetic environment due to fixed RF transmitters, an electromagnetic site

survey should be considered. If the measured field strength in the location in which the

CPRmeter is used exceeds the applicable RF compliance level above, the CPRmeter

should be observed to verify normal operation. If abnormal performance is observed,

additional measures may be necessary, such as re-orienting or relocating the CPRmeter.

Recommended separation distances between portable and mobile

RF communications equipment and the CPRmeter

The CPRmeter is intended for use in an electromagnetic environment in which

radiated RF disturbances are controlled. The customer or the user of the CPRmeter

can help prevent electromagnetic interference by maintaining a minimum distance

between portable and mobile RF communications equipment (transmitters) and the

CPRmeter as recommended below, according to the maximum output power of the

communications equipment.

Electromagnetic Emmissions

Rated maximum output

Separation distance according to frequency of

power of transmitter (W) transmitter (M)

80 MHz to 800 MHz

800 MHz to 2.5 GHz

d= 1.2

√P

d= 2.3

√P

0,23

0,72

2,3

7,28

23

0,12

0,38

1,2

3,8

12

0.01

0.1

1

10

100

For transmitters rated at a maximum output power not listed above, the

recommended separation distance d in meters (m) can be determined using the

equation applicable to the frequency of the transmitter, where P is the maximum

output power rating of the transmitter in watts (W) according to the transmitter

manufacturer.

NOTE: 1. At 80 MHz and 800 MHz, the separation distance for the higher frequency range

applies.

NOTE 2. The ISM (industrial, scientific and medial) bands between 150 kHz and 80 MHz

are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13, 567 MHz; 26,957 MHz to

27,283 MHz; and 40,660 MHz to 40,700 MHz.

NOTE 3. An additional factor of 10/3 is used in calculating the recommended separation

distance for transmitters in the ISM frequency bands between 150 kHz and 80 MHz

and in the frequency range 80 MHz to 2.5 GHz to

decrease the likelihood that mobile/portable communications equipment could cause

interference if it is inadvertently brought into patient areas.

NOTE 4. These guidelines may not apply in all situations. Electromagnetic propagation is

affected by absorption and reflection from structures, objects and people

NOTE 5. Transmitters/antennas of this power-level are most likely mounted on an

emergency vehicle chassis. The distances cited here are for open field. For an external

antenna, the separation distance is most likely shorter.

Advertising